Mainstream Capnograph Module

Part # SGMC

Easy to integrate, our capnography technology is available in inexpensive and reliable mainstream or sidestream versions for manufacturers of anaesthesia machines and ventilators.

SensoGas-MC mainstream capnograph module is an excellent plug and play alternative module for your CO2 gas analysis needs.

Thanks to its 2 types of airway adapter, it can be used on adult/children or neonates for continuous CO2 monitoring. As opposed to our competitors, our airway adapters are reusable, washable and disinfectable over 100 times, which represents a non negligible cost saving for a very close original purchasing cost per unit.

You may also be interested in our sidestream capnograph module for additional capnography applications on ventilators and anaesthesia machines.

capnorootmodule

  • Provide real-time data of EtCO2, InsCO2, Capnogram and Respiration Rate.
  • Real time transmission of module states: hardware state, software state and sensor state
  • Prevents water condensing with intelligent temperature control algorithm
  • Barometric pressure, temperature, anesthetic gases compensation
  • The small infrared probe will not increase breathing circuit load
  • Using high quality sensors, MTBF of 40000 hours
  • Equipped with DB9 electronic interface for easy connectivity
  • Independent electronic interface patch cord compatible with different monitor
  • Accuracy according to the requirements of ISO 21647:2004, YY0601-2009
  • For OEM version, can be delivered with compatible communication protocols
  • Electrical and Mechanical Specifications
• Voltage: +5V DC, offset voltage between ±10% of full range
• Power: Normal operation ≤ 1.5W; during Warm Up ≤ 2.5W
• Dimension: 43mm×32mm×26mm
• Weight: ≤ 35g (excluding cable)
• Cable Length: 3m ±0.2m

  • Performances, Range and Accuracy
CO2
• Range: 0 ~ 20 vol %
• Accuracy: from 0 ~ 10 vol %, ±(0.2vol% + 2% of reading); 
            from 11 ~ 20 vol %, ±(0.2vol% + 5% of reading)
Respiratory Rate
• Range: 0 ~ 150rpm
• Accuracy: from 0 ~ 70rmp, ±2rpm; above 70rmp, undefined
• Resolution: ±1rpm
Timings
• Rise time: ≤ 90ms
• System response time: ≤ 1s

  • Ratings
Working Conditions
• Temperature: 10 ~ 40°C (50 to 104°F)
• Humidity: 15 ~ 90%, non-condensation
• Altitude: 500 ~ 1200hPa
Storage Conditions
• Temperature: -20 ~ 70°C (4 to 158°F)
• Humidity: 15 ~ 95%, non-condensation
• Altitude: 400 ~ 1200hPa

Parameters SensoGas-MC Performances Competitor Performances
Measurement Range 0 ~ 20vol% (0 ~ 150mmHg) 0 ~ 15vol%
Measurement Accuracy 0~10 vol %, ±(0.2vol% + 2% of reading)
11~20 vol %, ±(0.2vol% + 5% of reading)
±(0.2vol%+ 2% of reading)

Click to download